This procedure applies to the 34461A, 34465A, and 34470A only (not applicable to the 34460A).
This procedure verifies AC and DC current on the 10 A terminals.
Values for the 34461A, 34465A, and 34470A DMMs | ||||||||
---|---|---|---|---|---|---|---|---|
Input | Error from Nominal | |||||||
Function | Current[1] | Frequency | Range | Quick Check |
24 hours | 90 days | 1 year | 2 years |
DCI | 5 A | N/A | 10 A | ± 3.5 mA | ± 7 mA | ± 8.5 mA | ||
DCI | 10 A | N/A | 10 A | Q |
± 16 mA |
± 23 mA | ± 26 mA | |
ACI | 10 A | 5 kHz | 10 A |
± 24 mA (34465A/70A) ± 29 mA (34461A) |
± 29 mA |
[1] If you are NOT using the Fluke 5725 for calibration, reduce the current to 2 A.